Free chlorine, total chlorine or monochloramine: choose the right result

Choose the residual your operating team needs, then match the sensor, reference check and monitoring point.

When specifying a chlorine analyser for a UK water installation, the first decision is which residual the team needs to follow. A free-chlorine result, a total-chlorine result and a monochloramine result provide different information.

PMA’s Halogen range provides separate sensor choices for these duties. Start with the water chemistry and the intended operating decision, then confirm the model, measurement range and installation.

Keep the three measurements distinct

Free chlorine describes the free residual. Total chlorine includes free and combined chlorine. Monochloramine is one specific combined-chlorine species, so its concentration should not automatically be treated as the entire total-chlorine result.

The PMA range identifies MP5 for free chlorine, MP Total for total chlorine and MP6 for monochloramine and free chlorine in suitable chloraminated water. Confirm the ordered variant and current manufacturer specification: product names and supported outputs need to be checked for the actual sensor supplied.

Name the residual on every record

Use the same residual name on the sensor specification, reference test and trend display. Free chlorine, total chlorine and monochloramine should not disappear behind a generic “chlorine” label.

PMA colleagues reviewing water-quality information

A representative location comes before an alarm

At a treatment works, decide whether the sample represents an outlet, a blend or another defined point. At a reservoir, consider the relationship between the inlet, outlet and an immersed sensor. Water travel and sample transport affect when a change appears on the trend.

Review an insertion, immersion or side-stream arrangement against the selected sensor and site conditions. Include pressure, mounting, isolation, drainage where relevant and maintenance access. The reservoir monitoring guide provides a starting point for that discussion.

Self-cleaning still needs verification

Halogen uses powered cleaning at the sensing surfaces and reagentless measurement. These features can reduce routine cleaning tasks, but the installation still needs reference checks, inspection and a documented operating routine.

Compare like with like: identify the residual measured by the reference procedure, collect a representative sample and account for timing. Record the sensor model, result, reference method, water conditions and any maintenance. Investigate differences before changing calibration or alarm settings.

Connect the result to the operating plan

For MP Total or another selected variant, agree the controller, outputs, valid-signal indication and responsibility for reviewing the trend. A residual reading is one part of the water-quality picture and needs the context of the site’s treatment and sampling procedures.

Send PMA the required residual, expected range, pH, conductivity, temperature, pressure and monitoring location. We can review the sensor and installation together, with suitable reference checks and support for your operating team.

Technical references

Halogen self-cleaning chlorine monitoring · Halogen MP Total · Chloraminated-water monitoring